8. Replace a defective mains power unit immediately.
9. ATMOS LC 27 Battery: Charge the battery fully, see chapter "3.3 Charging the
battery" on page 17.
10. Operate the device only with the bacterial filter cartridge inserted and with the
single-patient secretion canister system connected. The device is delivered with the
bacterial filter cartridge and the single-patient secretion canister system already
inserted and connected. Always keep an extra bacterial filter cartridge on hand.
11. Remove the single-patient secretion canister as instructed in chapter "4.5.1
Removing the single-patient secretion canister system" on page 24.
12. Before using it for the first time, clean the single-patient secretion canister, see
chapter "5.0 Cleaning and disinfection" on page 25.
 Recommendation: Before beginning suction, fill a little water (around 10 - 20 ml)
into the single-patient secretion canister system. This will prevent secretions from
drying into crusts.
13. Place the single-patient secretion canister into the canister guides on the base as
per chapter "3.6.1 Connecting the single-patient secretion canister system" on page
18.
14. Connect the suction hose.
15. Attach the suction hose to the single-patient secretion canister lid.

3.3 Charging the battery

 The battery's charge level is shown by the LED battery charge indicator on the
control panel.
 As the integrated battery is not automatically kept at fully operational level, you will
need to check the battery charge level regularly and, when necessary, have service
personnel replace it. The battery may be replaced by authorised and trained service
personnel only. You can check the battery status by switching on the device.
 The battery must be fully charged prior to first use.
1. Charge the battery at the latest when the battery charge indicator's red LED lights
up.
2. Only use the enclosed mains power unit (REF 011.1363.0).
3. Please observe the notes in chapter "6.3 Handling batteries" on page 30. Try to
avoid charging the battery only briefly, as this can damage the battery.
During battery recharging full suction capacity of the device is still available.
If the battery is fully discharged, you can operate the device using the mains power unit.
Understanding the battery charge
level:
Two green LEDs: > 60%
One green LED: 20 - 60%
One red LED: < 20%
During charging:
While the battery is charging, both green
LEDs are lit.