Symbol
Definition
Battery charging status
Key lock activated
Key lock deactivated
Leakage display. Shows leak tightness and/or leakage of the system using the
colours green, yellow and red. (Only visible during therapy)
Alert is suppressed. If the cause for the alert is remedied, the symbol will
automatically disappear.
USB stick connected
3.3

Prepare and assemble vacuum unit

3.3.1

Charging the battery

Prior to first start up of the ATMOS
the battery recharging unit supplied by ATMOS should be used. The vacuum unit should be
recharged in a cool place without direct solar irradiation. At ambient temperatures above 25°C
the charging time could be prolonged drastically. Improper use of the device may lead to serious
damage. Defects which occur due to improper handling of the device are not covered by the
guarantee.
• Plug the mains plug into the socket. The vacuum unit then displays the battery charging
status on the screen. The
• When the battery is fully charged (the
plug from the socket.
• After that remove the charging plug from the charging socket of the vacuum unit.
If the charging status of the battery is low, then the vacuum unit will raise an alert and the
screen with display a warning window (see chapter Warning messages). If the battery is too
low for further operation the vacuum unit switches off automatically.
S 042 NPWT the battery must be fully charged. Only
®
• Connect the plug of the battery recharging
unit to the charging connector
vacuum unit.
• Connect the battery recharging unit with
the supplied country specific power cable.
symbol changes.
symbol no longer changes) remove the mains
of the