Swann AMI Hydrogen QED Operator's Manual page 37

Table of Contents

Advertisement

AMI Hydrogen QED
Maintenance
A-96.250.791 / 040219
3 Press [Enter] to start the calibration and follow the dialog on the
display.
Calibration
Close regulating valve
to turn off sample flow.
<Enter> to continue
Calibration
Take sensor out of
flow cell and dry
membrane and sensor
<Enter> to continue
Calibration
Connect flow cell to a
source of hydrogen and
flush with hydrogen.
Dry flow cell inside.
<Enter> to continue
Calibration
Remount sensor and let
hydrogen stream through
flow cell gently until
value is accepted.
<Enter> to continue
Calibration
Saturation
Sat. Current
Progress
Calibration
Saturation
Sat. Current
<Enter> to save
If the calibration fails again, the hydrogen sensor has to replaced
by a new one.
4 Unscrew and remove the thread-
3.1.5
ed sleeve [A].
5 Remove the hydrogen sensor [B]
from the flow cell.
6 Dry the sensor membrane and the
3.1.5
flow cell with a soft paper tissue.
7 Flush the flow cell with hydrogen.
3.1.5
8 Stop hydrogen flow and dry the
flow cell.
9 Re-insert the hydrogen sensor into
the flow cell.
Make sure that the sensor
membrane is dry.
10 Start the hydrogen flow.
3.1.5
11 Press [Enter] to start the calibra-
tion measurement.
The saturation should reach 100%,
3.1.1
the saturation current should be about
xx.x%
2.0 A to 4.5 A. If the measuring val-
x.xx
A
ues are not stable during the measur-
ing period, the calibration will be
discarded.
If this is the case, check and if neces-
3.1.1
sary correct your measurement ar-
xx.x %
rangement and try again.
x.xx
A
If the calibration was successful press
[Enter] to save.
35