Technical Data - Swann AMI Hydrogen QED Operator's Manual

Table of Contents

Advertisement

AMI Hydrogen QED
Product Description
2.2.
Power Supply
Electronics
housing
Sample
requirements
Flow cell and
connection
Measuring
range
Accuracy
Reprodu-
cibility
A-96.250.791 / 040219

Technical Data

Voltage:
Power consumption:
Aluminium with a protection degree of IP 66 / NEMA 4X
Ambient temperature:
Limit range of operation:
Storage and transport:
Humidity:
Display:
Flow rate:
Temperature:
Inlet pressure:
Outlet pressure:
Flow cell made of acrylic glass with built-in flow adjustment valve
and digital sample flow meter
Sample inlet:
Sample outlet:
Range
0.01–9.99 ppb
10.0–99.9
100–800 ppb
0–50% saturation
±5% of measured value or ± 0.5 ppb
±1% of measured value or ± 0.5 ppb
100–240 VAC (± 10%)
50/60 Hz (± 5%)
or 24 VDC (± 10%)
max. 30 VA
-10 to +50 °C
-25 to +65 °C
-30 to +85 °C
10–90% rel., non condensing
backlit LCD, 75 x 45 mm
6 to 20 l/h
up to 45 °C
0.2 to 1 bar
pressure free
1/4" Swagelok tube adapter
flexible tube 8x6 mm
Resolution
0.01 ppb
0.1 ppb
1 ppb
0.1% Saturation
11