AMI Hydrogen QED
Program List and Explanations
5.3.2.6
5.3.2.7
5.3.4
5.3.4.1
5.3.4.2
5.3.4.3
A-96.250.791 / 040219
Signal Outputs: Select operating mode of the signal output:
Cont.:
Signal outputs continue to issue the measured value.
Hold:
Signal outputs hold the last valid measured value.
Measurement is interrupted. Errors, except fatal errors,
are not issued.
Off:
Signal outputs are switched off (set to 0 or 4 mA).
Errors, except fatal errors, are not issued.
Output/Control: Select operating mode of the controller output:
Cont.:
Controller continues normally.
Hold:
Controller continues based on the last valid value.
Off:
Controller is switched off.
Input: The functions of the relays and signal outputs can be
defined depending on the position of the input contact, i.e. no
function, closed or open.
Active: Define when the input should be active:
No:
Input is never active.
When closed
Input is active if the input relay is closed
When open:
Input is active if the input relay is open
Signal Outputs: Select the operation mode of the signal outputs
when the relay is active:
Cont.:
Signal outputs continue to issue the measured
value.
Hold:
Signal outputs issue the last valid measured value.
Measurement is interrupted. Errors, except fatal
errors, are not issued.
Off:
Set to 0 or 4 mA respectively. Errors, except fatal
errors, are not issued.
Output/Control: (relay or signal output):
Cont.:
Controller continues normally.
Hold:
Controller continues on the last valid value.
Off:
Controller is switched off.