Fault:
No:
No message is issued in pending error list and the
alarm relay does not close when input is active.
Message E024 is stored in the message list.
Yes:
Message E024 is issued and stored in the mes-
sage list. The Alarm relay closes when input is
active.
Delay: Time which the instrument waits, after the input is
deactivated, before returning to normal operation.
Range: 0–6'000 Sec
Language: Set the desired language.
Available settings: German /English/French/Spanish
Set defaults: Reset the instrument to factory default values in three
different ways:
 Calibration: Sets calibration values back to default. All other
values are kept in memory.
 In parts: Communication parameters are kept in memory. All
other values are set back to default values.
 Completely: Sets back all values including communication
parameters.
Load Firmware: Firmware updates should be done by instructed
service personnel only.
Password: Select a password different from 0000 to prevent
unauthorized access to the menus "Messages", "Maintenance",
"Operation" and "Installation".
Each menu may be protected by a different password.
If you forgot the passwords, contact the closest SWAN
representative.
Sample ID: Identify the process value with any meaning full text,
such as KKS number.
Line Break Detection: Define if message E028 should be issued in
case of a line break on signal output 1 or 2.
Choose between <Yes> or <No>.
A-96.250.791 / 040219