Maintenance Of The Faraday Electrode - Swann AMI Hydrogen QED Operator's Manual

Table of Contents

Advertisement

AMI Hydrogen QED
Maintenance
6.4
38

Maintenance of the Faraday Electrode

A
B
C
D
E
F
G
1 Switch off the instrument and close the flow regulating valve.
2 Open the Faraday control unit.
3 Disconnect and remove the cable from the Faraday control unit.
4 Unscrew and remove the fixing sleeve (A).
5 Remove the Faraday electrode from the flow cell, do not pull on
the cable.
6 Remove the washer (C) and the o-ring [D] from the electrode
body (B).
7 Unscrew the electrode tip containing the hollow electrode (G).
A
Fixing sleeve
B
Electrode body
C
Washer
D
O-ring
E
Inner electrode
F
O-ring
G
Hollow electrode
A-96.250.791 / 040219