AMI Hydrogen QED
Maintenance
6.3.3
Automatic
verification
Manual
verification
36
Faraday Verification
The Faraday verification works only for hydrogen concentrations
below 40–50 ppb. If automatic Faraday verification is enabled, a
periodic check of the system is performed. A manual verification
can be started for test purposes.
Per default the instrument performs an automatic Faraday verifica-
tion every 3 hours. To change the settings for automatic verification
navigate to menu <Operation>/<Faraday Parameter>, see menu
4.1.3, p. 58
for details.
Possible settings are:
off
interval
daily
weekly
To start a manual verification:
1 Navigate to menu 3.2.2 <Maintenance>/<Service>/<Faraday
Verification>.
Faraday Verification
Current Value
Faraday Conc.
12.85
Progress
<Enter> to stop
Faraday Verification
Efficiency
Faraday Conc.
12.85
Progress
Done
Results are saved in the Verification history menu 2.2.1.5
If the Faraday efficiency is below 50%, message E018 is displayed
and the Faraday electrode needs to be cleaned, see
of the Faraday Electrode, p.
2 Press [Enter] to start the Faraday
3.3
Verification.
ppb
1.62
The verification then starts
ppb
immediately.
3 Press [Enter] to confirm the Fara-
3.3
day Verification.
91.5 %
ppb
38.
Maintenance
A-96.250.791 / 040219