Faraday Verification - Swann AMI Hydrogen QED Operator's Manual

Table of Contents

Advertisement

AMI Hydrogen QED
Maintenance
6.3.3
Automatic
verification
Manual
verification
36

Faraday Verification

The Faraday verification works only for hydrogen concentrations
below 40–50 ppb. If automatic Faraday verification is enabled, a
periodic check of the system is performed. A manual verification
can be started for test purposes.
Per default the instrument performs an automatic Faraday verifica-
tion every 3 hours. To change the settings for automatic verification
navigate to menu <Operation>/<Faraday Parameter>, see menu
4.1.3, p. 58
for details.
Possible settings are:
 off
 interval
 daily
 weekly
To start a manual verification:
1 Navigate to menu 3.2.2 <Maintenance>/<Service>/<Faraday
Verification>.
Faraday Verification
Current Value
Faraday Conc.
12.85
Progress
<Enter> to stop
Faraday Verification
Efficiency
Faraday Conc.
12.85
Progress
Done
Results are saved in the Verification history menu 2.2.1.5
If the Faraday efficiency is below 50%, message E018 is displayed
and the Faraday electrode needs to be cleaned, see
of the Faraday Electrode, p.
2 Press [Enter] to start the Faraday
3.3
Verification.
ppb
1.62
The verification then starts
ppb
immediately.
3 Press [Enter] to confirm the Fara-
3.3
day Verification.
91.5 %
ppb
38.
Maintenance
A-96.250.791 / 040219