Swann AMI Hydrogen QED Operator's Manual page 77

Table of Contents

Advertisement

AMI Hydrogen QED
Default Values
Relay 1
Relay 2
A-96.250.791 / 040219
Sample Flow, Alarm low: ........................................................6.0 l/h
Sample Temp., Alarm High: ..................................................... 50 °C
Sample Temp., Alarm Low: ........................................................ 0 °C
Alarm Saturation; Alarm high.................................................. 120 %
Alarm Saturation; Alarm low .................................................... 0.0 %
Alarm Saturation; Hysteresis ...................................................... 2 %
Alarm Saturation; Delay............................................................... 5 s
Case temp. high: ..................................................................... 65 °C
Case temp. low:......................................................................... 0 °C
Function:.......................................................................... limit upper
Parameter:........................................................................ Hydrogen
Setpoint: .......................................................................... 10.00 ppm
Hysteresis:........................................................................... 100 ppb
Delay: ........................................................................................ 30 s
Function:.......................................................................... limit upper
Parameter:.................................................................... Temperature
Setpoint: .................................................................................. 50 °C
Hysteresis:................................................................................. 1 °C
Delay: ........................................................................................ 30 s
If Function = Control upw. or dnw:
Parameter:.................................................................... Meas. Value
Settings: Actuator: .......................................................... Frequency
Settings: Pulse Frequency: ..............................................120/min
Settings: Control Parameters: Setpoint:....................... 10.00 ppm
Settings: Control Parameters: P-band: ............................ 100 ppb
Settings: Control Parameters: Reset time:............................... 0 s
Settings: Control Parameters: Derivative Time: ....................... 0 s
Settings: Control Parameters: Control Timeout:................... 0 min
Settings: Act. Time prop.: Cycle time: ................................... 60 s
Settings: Act. Time prop.: Response time: ............................ 10 s
Settings: Act. Motor valve: Run time: .................................... 60 s
Settings: Act. Motor valve: Neutral zone: ................................ 5%
If Function = Timer:
Mode:.................................................................................... Interval
Interval: ................................................................................ 1 min
Mode: ........................................................................................daily
Start time:........................................................................ 00.00.00
Mode:..................................................................................... weekly
Calendar; Start time: ....................................................... 00.00.00
Calendar; Monday to Sunday:.................................................. Off
Run time: ................................................................................... 10 s
Delay: .......................................................................................... 5 s
75