Monitor AMI Hydrogen QED
B
AMI Inspector Hydrogen
C
Reference flow cell
D
On-line flow cell
1 Stop sample flow to the monitor AMI Hydrogen QED by closing
the appropriate valve, e.g. back pressure regulator, sample
preparation or flow regulating valve at flow cell.
2 Connect sample line of the monitor AMI Hydrogen QED [A] with
the sample inlet of the reference instrument AMI Inspector Hy-
drogen [B]. Use the supplied tube.
3 Connect sample outlet of the reference instrument AMI Inspec-
tor Hydrogen to the sample outlet funnel of the monitor.
4 Switch on the AMI Inspector Hydrogen. Open the flow regulat-
ing valve and regulate the sample flow to 10 l/h. The actual flow
is shown on the transmitter.
A
B
C
D
E
F
G
E
Sample outlet
F
Sample inlet
G
T-fitting