Swann AMI Hydrogen QED Operator's Manual page 50

Table of Contents

Advertisement

AMI Hydrogen QED
Troubleshooting
Error
E013
Case Temp. high
E014
Case Temp. low
E017
Control Timeout
E018
Faraday Efficiency
E019
Quality Assurance
E024
Input active
E026
IC LM75
E028
Signal output open
E030
EEProm Frontend
E031
Calibration Recout
E032
Wrong Frontend
E033
Power-on
E034
Power-down
48
Description
Corrective action
– check case/environment temperature
– check programmed value,
see
5.3.1.5.1, p. 67
– check case/environment temperature
– check programmed value,
see
5.3.1.5.2, p. 67
– check control device or programming in
Installation, Relay contact, Relay 1/2
see
5.3.2 and 5.3.3, p. 67
– clean Faraday electrode, see
Maintenance of the Faraday Electrode,
p. 38
– perform QA Procedure using a
reference instrument, e.g. AMI Inspector
– see If Fault Yes is programmed in Menu
see
5.3.4, p. 71
– call service
– check wiring on signal outputs 1 and 2
– call service
– call service
– call service
– none, normal status
– none, normal status
A-96.250.791 / 040219