or full bottle. This alarm output stays on
as long as the run time display reads
PGM: DONE.
• Recycle: The collection of incomplete
samples by repeating the sampling cycle
as many as four times. To enable or
disable it.
• Temperature Error: The temperature
has been above its set temperature for
more than 30 minutes.
• Failed Sample: This event occurs when
the sampler failed to reach the
liquid-sensor electrodes. This may occur
if the intake line is plugged or the
vacuum time may need to be increased.
• Indexing Error: This event occurs when
the turntable has an indexing failure.
This condition must be corrected before
the sampling program can resume. This
alarm output stays on until the condition
is corrected.
(Alarm- 7 pin output connector using 2 red
wires pins 5 & 6.)
• Run Status: This event occurs each time
a sample is taken. The output stays on
until the sample event is completed.
(2 blue wires pins 3&4)
• Float Start or Float Start/Stop:
*Float Start- If the FLT S
selected, the program will start with a
closed contact and the program will continue
Section 3 Programming
setting is
TART
3-11