Platinum/Optima CVE Refrigerated Sampler
Sample Seconds: The sample size is set with
the adjustable intake tube. Set a sampling time
duration that is long enough to allow time for
the sample to fill the sample chamber and
touch the liquid sensing electrodes. Test this
and increase the time if no sample enters the
chamber. Press F3 to accept
.
Postpurge: Enter the number of seconds
(10-99) for the system to purge the intake after
the sample is taken. Press F3 to accept.
Num Rinses: Enter the number of times (0-4)
the intake line should be rinsed, prior to each
sample. Press F3 to accept. This option is to
precondition the sample line before taking a
sample.
Pressure Time: This menu item displays only
if the Number of Rinses is not zero. Type in the
number of seconds (0-99) each rinse should
pressurize the sampler chamber.
Vacuum Time: Enter the number of seconds
(0-99) each rinse should generate suction. The
time should be short enough that the sample
does not enter the sample chamber during the
rinse. Press F3 to accept.
Recycle: Use up and down arrows to enable or
disable the sample retry feature, or press F3 to
accept the setting. When the recycle feature is
enabled, the sampler makes a second attempt
to take the sample, if it failed to collect a
sample on the first try.
Consecutive SMPL: To draw just one sample
per interval, type 1 and press F3 to accept.