System Description

Specifications

Table 1.2 Run Parameter Icons (Continued)
Specifications
Only values with tolerances or limits are guaranteed data. Values without tolerances are informative data,
without guarantee.
Table 1.3 Specifications
Specification
Set Speed
Set RCF
Speed
Speed Display
Speed Accuracy
Set Time
Time Display
Time
Set Temperature
Temperature Display
Temperature Accuracy
Temperature
Over Temperature
Shutdown
Acceleration
Acceleration Profiles
Deceleration
Deceleration Profiles
1-10
SOUND OFF — In the Menu screen, press to turn the sounds on.
SERVICE — For Beckman Coulter Service Use only.
INFORMATION — In the Menu screen, press to see instrument and software
information.
200 to 10,200 RPM
in 10 RPM increments
10 to 11,420 x g
in 10 x g increments
actual rotor speed in 10 RPM increments or actual RCF in 10 x g
increments
±25 rpm of Set Speed from 200 to 10,200 RPM
1 minute to 99 hours 59 minutes or continuous (hold)
Timed Run: indicates run time remaining (HH:MM:SS)
Hold Run: indicates elapsed time (HH:MM:SS)
Pulse Run: indicates elapsed time (HH:MM:SS)
N/A
a
>55°C
b
10 acceleration rates, including maximum torque
11 deceleration rates, including maximum torque and no braking
J-15 Ventilated
200 to 10,200 RPM
in 10 RPM increments
10 to 11,420 x g
in 10 x g increments
–10 to +40°C in 1°C increments
Chamber temperature in 1°C
increments
±2°C of Chamber temperature
(after equilibration); applies to 4 to
25°C temperature range
>55°C
J-15R Refrigerated
a
PN B80286AC