Beckman Coulter Avanti J-15 Series Instructions For Use Manual page 33

Table of Contents

Advertisement

Table 1.3 Specifications (Continued)
Specification
Width
Dimensions
Depth
Height
Weight
Weight, not including rotor
Sides
Ventilation
c
clearances
Rear
Top Surface
Finishes
Front Surface
Door
Electrical Requirements
Electrical
Electrical Supply
Installation (overvoltage)
category
Noise output (1 m in front of
instrument, 1.5 m above the
floor with JA-10.100 rotor at
10,200 rpm)
Ambient Temperature Range 10 to 31°C
Humidity
Environmental
Refrigerant
Maximum heat dissipation
into room under steady-state
conditions
Pollution degree
Altitude
a. Display Temperature vs. Chamber Temperature: To reach temperatures above ambient, the centrifuge is dependent on the frictional heat
generated inside the chamber during operation. At low run speeds or low ambient temperatures, the centrifuge may not be able to achieve
some higher temperatures. At high run speeds or high ambient temperatures, the centrifuge may not be able to achieve some lower
temperatures.
b. If the system reaches this temperature, it will issue a diagnostic and shut down using maximum brake.
c. When operating the J-15 Centrifuge, maintain a 30 cm (12 in.) clearance envelope around the centrifuge under the following circumstances:
1) Run duration exceeds 60 minutes or 2) Ambient temperature exceeds 25°C.
d. Normally only nonconductive pollution occurs; occasionally however, a temporary conductivity caused by condensation must be expected.
PN B80286AC
J-15 Ventilated
55.6 cm (21.9 in.)
74.9 cm (29.5 in)
36.8 cm (14.5 in)
93 kg (205 lbs)
7.6 cm (3.0 in.)
7.6 cm (3.0 in.)
Painted steel
Uncoated plastic
Painted Aluminum and plastic
100 VAC, 12A, 50/60 Hz
120 VAC, 10A, 50/60 Hz
200-230 VAC, 6A, 50/60-Hz
Class 1
II
61 dBA
80% non-condensing
N/A
4095 Btu/h (1.2kW)
d
2
up to 2000 meters
System Description
Specifications
J-15R Refrigerated
75.6 cm (29.8 in)
70.3 cm (27.7 in)
36.8 cm (14.5 in)
120 kg (265 lbs)
120 VAC, 12A, 60 Hz
200-230 VAC, 8A, 50 Hz
208-230 VAC, 9A, 60 Hz
58 dBA
10 to 35°C
80% non-condensing
R404A
120V: 4913 Btu/h (1.44kW)
200-230V: 6551 Btu/h (1.92 kW)
1
1-11