Available Rotors - Beckman Coulter Avanti J-15 Series Instructions For Use Manual

Table of Contents

Advertisement

System Description

Available Rotors

Available Rotors
The following Beckman Coulter rotors can be used in the Avanti J-15 and J-15R centrifuges. The
rotors are described in individual manuals that accompany each rotor.
Table 1.4 Available Rotors
Rotor Profile
Description
JA-10.100
Fixed Angle
r
= 98.0 mm
max
r
= 35.0 mm
min
JS-4.750
Swinging Bucket
Tube-and-bottle
buckets
r
= 207.8 mm
max
JS-4.750
Swinging Bucket
Multiwell-plate
carriers,
r
= 183.2 mm
max
a. Maximum speeds are based on a solution density of 1.2 g/mL. At upper temperature and humidity ambient conditions, swinging bucket rotor
speed may require reduction.
b. Relative Centrifugal Field (RCF) is the ratio of the centrifugal acceleration at a specified radius and speed (r
of gravity (g) according to the following formula: RCF = r
second (2  RPM/60), and g is the standard acceleration of gravity (9807 mm/s
1-12
a
Max RPM
J-15
J-15R
10,200
10,200
4550 for
120V unit.
4550
4750 for
200-230V
unit
1 to 3
1 to 3
stacked
stacked
plates:
plates
4350
4450
4
stacked
4 stacked
plates:
plates: 2700
2700
2
/g—where r is the radius in millimeters,  is the angular velocity in radians per
b
Max RCF
( g)
Number
at r
of Tubes 
max
Nominal
J-15
J-15R
Capacity
11420
11420 × g
6 × 100 mL
× g
4820 × g
for
120V unit
4820
4 × 750
5250 × g
× g
for
200- 230V
unit
3880
4 × 3 × 96
4060 × g
× g
1500
4 × 4 × 96
1500 × g
× g
2
) to the standard acceleration
2
). After substitution: RCF = 1.12 r (RPM/1000)
Rotor
Manual
Number
B80290
B80289
mL
B80289
mL
B80289
mL
2
.
PN B80286AC

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents