INDICATOR TROUBLESHOOTING TABLE

The following is a troubleshooting table for the AED indicators.
VIEW

SCHEDULED MAINTENANCE

DAILY MAINTENANCE
Check the STATUS INDICATOR to ensure that it is GREEN. When the indicator is GREEN, the
Responder AED Pro is ready for a rescue. If the indicator is RED, refer to the Troubleshooting Table in this
chapter.
MONTHLY MAINTENANCE
1.
Open the AED lid.
2.
Wait for the AED to indicate status
3.
Observe the change of the STATUS INDICATOR to RED. After less than 5 seconds, verify that the
STATUS INDICATOR returns to GREEN.
4.
Observe the expiration date on the pads.
5.
Listen for the voice prompts.
6.
Close the lid and confirm that STATUS INDICATOR remains GREEN.
Page 22 of 33
SYMPTOM
Red SERVICE indicator (LED) is
indicated on the screen.
Red PADS indicator (LED) is indicated
on the screen.
The LAST BATTERY indicator (LED) is
red and flashing.
STATUS INDICATOR is RED, and no
other indicators on the diagnostic panel
are lit.
© 2007 General Electric Company
SOLUTION
Maintenance by authorized service personnel
is required. Call Customer Service or your
local distributor.
Connect the pads or replace with a new pair.
The battery is low. Replace with a new
battery.
The battery power is completely depleted.
Replace with a new battery. If STATUS
INDICATOR remains RED, refer to the
Responder AED Pro for maintenance.
Call Customer Service or your local
distributor.