Description
Set Sensitivity
Read Setpoint States
Read assigned setpoints #aa04cn
Set Setpoint
OFF/ON/Auto
B-6
XGS-600 Gauge Controller
Table B-1 Serial Command Set (Continued)
Command
#aa55cnxx.xx
Setpoint Commands
#aa03
#aa5Ehx
where h is setpoint
number 1-8 and x is
state:
0 = OFF (manual
override)
1 = ON (manual
override)
3 = Auto
(based on pressure)
Example: #005E83 sets
setpoint #8 to Auto.
Response from
XGS-600
>
>00hh
where hh is a hex value
where each bit
represents
setpoints 1 - 8 and
value 0=Off, 1=On
>00hh
where hh is a hex value
where each bit
represents
setpoints 1 - 8 and
value 0=Not assigned,
1=Assigned
>
Notes
This sensitivity value
must always be in Torr
due to xx.xx format
restriction. Note
however the value
displayed on front
panel is in Torr,
mbar, or Pascal.
Least significant digit
will be ignored.
Note: Will overwrite
any factory setting.
Setpoint numbers
correspond to the
numbers on the
Setpoint screen.

Advertisement

Table of Contents
loading

Table of Contents