Set Point Screen - Varian XGS-600 Instruction Manual

Gauge controller
Table of Contents

Advertisement

Set Point Screen

To open this screen select SETPTS on the Main screen.
Set Point #
SETPT
ON
ON DELAY
XGS-600 Gauge Controller
Figure 3-3 Set Point Screen
This number corresponds to the sequence of eight rear panel Set
Point outputs. This is also the number that appears on the Main
screen if the setpoint is activated. These are fixed numbers and not
user changeable.
Select an available gauge in this list to assign to a Setpoint #. Only
two set points are allowed per sensor. If two setpoints are already
assigned to a sensor, that sensor does not appear in the list of
options for this field.
NONE is available as a choice. If selected, all pressures and delays
are set to zero and OFF.
Select a Setpoint On pressure value. If the sensor pressure falls
below this value, the rear panel output activates (voltage low) and
the Setpoint # appears on the Main screen next to that sensor. The
ON reading must be a lower pressure than the OFF reading or an
error (ERR) warning appears in the rightmost column. Once valid
settings are entered, ERR changes to AUTO.
See Appendix Appendix A. "XGS-600 Gauge Specifications" on
page A-1 for setpoint output specifications and setpoint connector
pin-outs.
If a delay is required between the time the sensor detects the
Setpoint ON pressure value and the outputs activates (rear panel,
Main screen indicator and serial communication), select a delay
time in seconds. Example: A setting of 1.2 results in a delay of
approximately 1.2 seconds.
3-7