Img Board Specifications - Varian XGS-600 Instruction Manual

Gauge controller
Table of Contents

Advertisement

Item
Remote I/O Connector
Allowable Cable Length
IMG Board
The IMG board operates Varian IMG-100 and IMG-300 gauges and allows you to set gauge
sensitivity value, and gas correction factor. The UHV range is standard, as is remote I/O
control of emission and 0 to 10 V analog output.
Item
Supported Gauge Heads
Displayed Measurement Range*
Sensitivity Input Range
Gas Correction Factor Range
Analog Output
Error Detection
Gauge Connector
Remote I/O Connector
Allowable Cable Length
XGS-600 Gauge Controller
Table A-3 Convection Board Specifications
6 pin Firewire type
Grounding the input pin activates that function
The Status Output goes hi during an error.
Note: the analog outputs for CNV1 and CNV2 are also
available at pin 7 of their respective Dsub connectors.
up to 300' using Varian TC or ConvecTorr cable
Table A-4 IMG Board Specifications
❑ Varian IMG-100, IMG-300
❑ 5.0E-12 Torr to 2 mTorr (IMG-300)
❑ 5.0E-12 Torr to 5 mTorr (IMG-100)
* This represents the ability of the controller to display pressure
values and not the capability of the gauge heads to accurately
measure pressure.
00.1 to 99.9
0.1 to 9.9
❑ Range: 0-10 VDC
❑ Slope: 1 V per decade log
❑ Characteristic: output voltage = Log (Pressure) + 11
❑ Output when gauge OFF or Error = 10 V
❑ Load Impedance: >10 kOhms
❑ Over Temperature
❑ Over Pressure (2 mT or 5 mT dependent on tube type selected)
❑ SHV
❑ 6 pin Firewire type
Grounding the input pin activates that function
The Status Output goes HI during an error.
❑ up to 300' using RG59U
Description
Description
A-5

Advertisement

Table of Contents
loading

Table of Contents