Sensor ID
User Label
Sensor Model
3-10
XGS-600 Gauge Controller
Use this field to select the sensor to view its setup parameters.
Displays the unique system-assigned gauge identification. Scroll
through this list to select a sensor and the remainder of fields on the
screen populate with its present settings.
Possible sensor ID's are:
HFIGx for Hot Cathode Ion Gauge boards using BA or UHV
sensors.
CNVx for Convection boards (TC/ConvecTorr)
IMGx for IMG Gauge boards
where x indicates the number of the transducer attached. The
number is system-assigned based on the sequence of board
installation, left to right (when viewed from the front). Convection
boards have two sensors per board. When more than 4 convection
boards are installed, the 10, 11, and 12 channels are designated
CNVA, CNVB, and CNVC, respectively.
Use this field to enter your label for this sensor that appears on all
other screens and serial communication. This field has a maximum
of five (5) characters and any alphanumeric character is available
for use.
User Labels that begin with HFIG, CNV or IMG are not allowed to
avoid confusion with Sensor ID's.
If no User Label is selected, the default is the Sensor ID. When a
System Reset is initiated, this label is reset to the sensor ID.
Use this field to configure the particular model of sensor attached
to a board.
NOTE
If Sensor Model is changed, all fields resort to default values.