System Setup Screen

Use this screen to set values for overall system operation.
To open this screen select SYSTEM SETUP on the Sensor Setup screen.
UNITS
SETUP LOCKOUT
AUTO FILAMENT
ADVANCE (HFIG
Boards)
3-14
XGS-600 Gauge Controller
Figure 3-6 System Setup Screen
Select: Torr (default), Pascal or mbar. Once the units are changed,
the values throughout the system are updated to the new units.
Toggle this to ON to make all fields on the SYSTEM SETUP and
SENSOR SETUP screen display-only (except for this field). This
provides a layer of security to prevent inadvertent changes.
For two-filament HFIG sensors, if Auto Filament Advance is
selected, Filament 2 is automatically engaged if Filament 1 has an
open filament error (NO FIL1). FIL2 ON then appears on the main
screen. Auto filament advance can occur when emission is first
turned on or during gauge operation. Auto advance does not
advance from Filament 2 to Filament 1. If an open filament error
occurs on Filament 2, instead the NO FIL2 error appears and
emission is turned off.