c. Locate the rear panel portion of the board into its slot, press down the tab and
d. Insert the yellow HFIG plug into the connector (HFIG boards only). See
e. Inspect the installation to ensure the board is seated properly and not touching
6. Replace the covers:
a. Push the cover straight in under the lip at the XGS front bezel.
b. Replace the two Philips head screws.
c. The cover must be flat and fully seated. It may be necessary to adjust the boards
7. Plug the unit back in and turn on the power switch.
4-4
XGS-600 Gauge Controller
exert gentle pressure until it clicks into place.
CAUTION
Do not force the board down. If too much pressure is required
the connector pins are not aligned.
Figure 4-3. This connector is keyed.
any adjacent boards.
and filler plates slightly for the cover to fully engage. The cover should drop on
front of the filler plate.
NOTE
Any board-specific settings must be reprogrammed.
Most Sensor setup and Set point user settings stay with the
board. This means if the removed board is used elsewhere,
most settings will be saved. However, if a new board is
installed, you will have to reprogram the settings.