Varian XGS-600 Instruction Manual page 61

Gauge controller
Table of Contents

Advertisement

Description
Read Setpoint
OFF/ON/Auto
Set Setpoint On
pressure level
Set Setpoint ON Delay
time
Set Setpoint Off
pressure
level
Set Setpoint Off Delay
time
Read Setpoint On
pressure level
Read Setpoint ON
Delay time
Read Setpoint Off
pressure level
XGS-600 Gauge Controller
Table B-1 Serial Command Set (Continued)
Command
#aa5Fh
where h is setpoint
number 1-8
#aa6hcnx.xxxE-xx
where h is setpoint
number 1-8
#aaChx.x
where h is setpoint
number 1-8 and x.x is
delay in seconds.
C is part of the hex
command number.
Example: #00C31.2 sets
on delay to 1.2 seconds
for setpoint # 3
#aa7hcnx.xxxE-xx
where h is setpoint
number 1 - 8
#aaDhx.x
where h is setpoint
number 1-8 and x.x is
delay in seconds
#aa8h
where h is setpoint
number 1 - 8
#aaEh
where h is setpoint
number 1-8
#aa9h
where h is setpoint
number 1 - 8
Response from
XGS-600
>x
where x is state:
0 = OFF (manual
override)
1 = ON (manual
override)
3 = Auto (based on
pressure)
>
>
>
>
>x.xxxE-xx
>x.x
where x.x is delay in
seconds
>x.xxxE-xx
Notes
Setpoint must first be
assigned to a sensor
using setpoint screen or
command 6h.
Setpoint ON level must
be set before OFF level.
Setpoint must be
assigned to a sensor.
B-7