Varian XGS-600 Instruction Manual page 57

Gauge controller
Table of Contents

Advertisement

Description
Set pressure units to
Torr
Set pressure units to
mBar
Set pressure units to
Pascal
Read pressure units
Assign User Label to
Sensor
Read User Label of
Sensor
Setup Lockout OFF
Setup Lockout ON
Read Setup Lockout
status
Set Auto-On
XGS-600 Gauge Controller
Table B-1 Serial Command Set (Continued)
Command
#aa10
#aa11
#aa12
#aa13
#aa14cnLABEL
where c sensor code
=I for ion gauges HFIG
and IMG
=T for CNV's
n = Sensor Count
counting TC's or ion
gauges from left to right
from the front panel
view.
LABEL= user-defined
1-5 characters.
#aa15cn
#aa20
#aa21
#aa22
#aaB0cn1cn2x.xE-xx
where cn1 is the Ion
Gauge and cn2 is the
TC.
Response from
XGS-600
>
>
>
>xx
where:
xx = 00 is Torr
xx = 01 is mBar
xx = 02 is Pascal
>
>LABEL
where LABEL is
user-defined 1-5
characters.
>
>
>xx
where:
xx = 00 is OFF
xx = 01 is ON
>
Notes
Label can be any 5 or
fewer characters
(AHZ, 0-9, or space),
except for IMGxx,
HFIGx and CNVxxx,
(where x is any
character) which are
reserved.
If no User Label is
assigned, defaults to
system-assigned Sensor
ID.
B-3

Advertisement

Table of Contents
loading

Table of Contents