  • Page 1 vacuum technologies MiniVac Controller MANUALE DI ISTRUZIONI BEDIENUNGSHANDBUCH Model 929-0190 NOTICE DE MODE D’EMPLOI Model 929-0290 (GB) INSTRUCTION MANUAL Model 929-0191 Model 929-0291 87-900-049-01 (E) APRIL 2006...
  • Page 2 MiniVac Controller...
  • Page 5: Table Of Contents

    ISTRUZIONI PER L’USO ....................1 GEBRAUCHSANLEITUNG ....................4 MODE D’EMPLOI .......................7 INSTRUCTIONS FOR USE ....................10 TECHNICAL INFORMATION ....................13 MINIVAC DESCRIPTION ........................13 CONTROLLER SPECIFICATIONS......................15 CONTROLLER OUTLINE ........................16 INSTALLATION ............................17 Line Voltage Change Over ........................17 Output Polarity Selection ........................17 Input/Output Interconnections ......................18 Pump Connection..........................18 OPERATION ............................19 Pressure Determination........................19 Power Failure ............................19...
  • Page 6: Istruzioni Per L'uso

    • temperatura: da -20 °C a +70 °C • umidità relativa: 0 - 95% (non condensante) Ogni controller giunge dalla Varian predisposto per una certa tensione di alimentazione: • il modello 929-0190 KINGS type High voltage connector (120 Vac, 50 – 60 Hz) •...
  • Page 7 ISTRUZIONI PER L’USO INSTALLAZIONE Comandi, Indicatori e Connettori del Controller Di seguito sono illustrati sia il pannello frontale che quello posteriore con le interconnessioni. PERICOLO! Per maggiori dettagli fare riferimento alla sezione "Technical Information". Il controller è fornito di un cavo di alimentazione a tre fili con una spina di tipo approvato a livello internazionale.
  • Page 8 In caso di guasto è possibile usufruire del servizio di esso la casa madre o un rivenditore, per avviare il riparazione Varian o del "Varian advance exchange processo di raccolta e smaltimento, dopo opportuna service", che permette di ottenere un controller verifica dei termini e condizioni contrattuali di vendita.
  • Page 9: Gebrauchsanleitung

    Geltung kommenden Bestimmungen übernimmt die werden. Es ist vollständig recyclebar und entspricht der Firma Varian keinerlei Haftung. Der MiniVac ist ein EG-Richtlinie 85/399 für Umweltschutz. Hochspannungsnetzgerät niedriger Leistung, das zur Versorgung von Ionenpumpen verwendet wird. In den folgenden Abschnitten sind alle erforderlichen Informationen für die Sicherheit des Bedieners bei der...
  • Page 10 GEBRAUCHSANLEITUNG INSTALLATION Steuerungen, Anzeigen und Anschlüsse des Controllers Nachstehend werden dass Bedienfeld des Controllers sowie die Anschlußfelder beschrieben. GEFAHR! Der Controller wird mit einem 3-adrigen Netzkabel Für weitere Einzelheiten siehe "Technical Information". geliefert, das mit einem den internationalen Normen entsprechenden Stecker ausgerüstet ist. Es sollte immer dieses Netzkabel benutzt werden, das an eine korrekt geerdete Steckdose anzuschließen ist, um Stromentladungen zu vermeiden.
  • Page 11 Haushaltsmüll entsorgt werden darf sondern einem Wartung. Alle Eingriffe dürfen nur von autorisiertem speziellen Sammelsystem zugeführt werden muss. Personal vorgenommen werden. Der Endabnehmer sollte daher den Lieferanten des Bei Defekten kann der Reparatur Service von Varian Geräts d.h. Muttergesellschaft oder oder der "Varian Advance Exchange Service"...
  • Page 12: Mode D'emploi

    Température: de - 20 °C à + 70 °C Emballage des Contrôleurs • Humidité relative: 0 - 95% (non condensante). Chaque contrôleur est fourni par Varian prééquipé pour une certaine tension d'alimentation: • Modèle 929-0190 KINGS type High voltage connector (120 Vac, 50 – 60 Hz) •...
  • Page 13 MODE D'EMPLOI INSTALLATION Commandes, Indicateurs et Connecteurs du Contrôleur La figure ci-après représente le tableau de commande du Contrôleur et les tableaux d'interconnexion. DANGER! Pour de plus amples détails, se reporter à la section Le contrôleur est doté d'un câble d'alimentation à trois "Technical Information".
  • Page 14 En cas de panne, il est possible de faire appel au fournisseur du dispositif, maison mère ou revendeur, service de réparation Varian ou au "Varian advance pour mettre en œuvre le processus de collecte et mise exchange service", qui permet d’obtenir un contrôler au rebut.
  • Page 15: Instructions For Use

    If this shows signs of damage which may have other additional information supplied by Varian before occurred during transport, contact your local sales operating the equipment. Varian will not be held office. When unpacking, ensure that the module is not responsible for any events occurring due to non- dropped or subjected to any form of impact.
  • Page 16 INSTRUCTIONS FOR USE INSTALLATION Controller Controls, Indicators and Connectors The following paragraph illustrates either the front and rear panel with its interconnections. WARNING! The controller is equipped with a 3-wire power cord and More details are contained in the appendix "Technical plug (internationally approved) for user's safety.
  • Page 17 When a fault has occurred it is possible to use the retailer, to initiate the collection and disposal process Varian repair service or the “Varian advance exchange after checking the contractual terms and conditions of service”, that allows to obtain a regenerated controller sale.
  • Page 18: Technical Information

    TECHNICAL INFORMATION MINIVAC DESCRIPTION The MiniVac controller is a quarter rack solid state Medium/larger ion pumps must first be started with an power supply with self protection features, which appropriate control unit and pumped to a sufficiently generates high voltage using high frequency switching low pressure where the current demand is below 8 mA.
  • Page 19 TECHNICAL INFORMATION The MiniVac controller front panel controls are shown The MiniVac controller rear panel controls are shown in in the following figure. the following figure. Mini Vac MiniVac Controller Rear Panel Controls MiniVac Front Panel Controls Controller power entry module consisting of mains High Load LED lights up when the pump current is fuses, voltage change over, mains socket.
  • Page 20: Controller Specifications

    TECHNICAL INFORMATION CONTROLLER SPECIFICATIONS Operation All type and size of Voltage VacIon pump Recorder Output: Input: - Rec Out 0 to 5 V (corresponding to 0 to 5 kV with 1 V - Voltage 100 Vac corresponding to 1 kV in (4 ranges) 120 Vac linear mode)
  • Page 21: Controller Outline

    TECHNICAL INFORMATION Voltage Vs Current Diagram CONTROLLER OUTLINE The outline dimensions of the MiniVac controller are shown in the following figure. MiniVac Outline Dimensions 87-900-049-01 (E)
  • Page 22: Installation

    TECHNICAL INFORMATION INSTALLATION Line Voltage Change Over If a change in line voltage operation is desired, proceed Inspect the controller for any shipping damage. as follows: • Models 929-0190 and 929-0191 are factory set for • Unplug the power cord from the controller rear 120 Vac operation panel socket.
  • Page 23: Input/Output Interconnections

    TECHNICAL INFORMATION Input/Output Interconnections All input/output signals are accessible at J001 The front panel REC.OUT connector is a duplicate of accessory connector. With the provided P001 mating this signal on a BNC connector. connector make connections with AWG 24 (0.2 mm ) - or smaller wire - to the pins Pin 2-6 Current recorder output 2 (pin 2 signal out,...
  • Page 24: Operation

    TECHNICAL INFORMATION OPERATION Pressure Determination Pressure at the inlet flange of a VacIon pump is Make all vacuum manifold and electrical connections proportional to the current drawn by the pump. and refer to the ion pump instruction manual prior to operating the MiniVac controller.
  • Page 25: Maintenance

    TECHNICAL INFORMATION MAINTENANCE Controller Spares and Accessories Replacement controllers are available on advance exchange basis through Varian service/sales Description Part number organization. Mating accessory connector 03-648487 P001 WARNING Fuse T 2A (5x20 mm) High voltage in the controller can cause severe injury or death.
  • Page 26 NOTE: If a product is received at Varian which is contaminated with a toxic or hazardous material that was not disclosed, the customer will be held responsible for all costs incurred to ensure the safe handling of the product, and is liable for any harm or injury to Varian employees as well as to any third party occurring as a result of exposure to toxic or hazardous materials present in the product.
  • Page 27 Request for Return FAILURE REPORT TURBO PUMPS and TURBOCONTROLLERS POSITION PARAMETERS Power: Rotational Speed: Does not start Noise Vertical Current: Inlet Pressure: Does not spin freely Vibrations Horizontal Temp 1: Foreline Pressure: Does not reach full speed Leak Upside-down Mechanical Contact Overtemperature Other: Temp 2:...
  • Page 28 Toll Free # 1 (800) 882 7426 Tel: 00 800 234 234 00 vtt.technical.support@varianinc.com China Mexico Varian Technologies - Beijing Varian, S. de R.L. de C.V. Rm 1648 Central Tower South Wing Concepcion Beistegui No 109 China Beijing Junefield Plaza Col Del Valle No.