Controller Specifications - Varian Turbo-V 2000 ICE Instruction Manual

Hide thumbs Also See for Turbo-V 2000 ICE:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 97

CONTROLLER SPECIFICATIONS

Input:
Voltage
Frequency
Power
Output for pump:
Voltage
Frequency
Start up power
Maximum output power:
Process gas Argon
Process gas Air
Operating temperature
Storage temperature
Circuit breaker
Fuse Heater Jacket
P1 optoisolator:
J2 semi specifications connector:
J7 output (fan, forepump coil, etc.)
J11 purge flow connector
J12 heater jacket connector
In compliance with norms
Auxiliary connectors
J1
J2
J7
J13
Interconnecting cables
Weight (both models)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
100, 120, 220, 240 Vac, 1-phase
50 - 60 Hz
1900 VA maximum
120 Vac nominal ±10%
550 Hz ±2%
700 W maximum
300 W
450 W
0 °C to +40 °C
-20 °C to +70 °C
10 A for 220 V input Voltage
10 A for 220 V input Voltage
T3.15A (slow blow)
inputs
Minimum ON 3 mA
outputs
24 Vdc, 60 mA
inputs
Minimum ON 3 mA
outputs
24 Vdc, 100 mA
120 Vac, 1A disregarding the mains
12 Vdc power supply output and signal input
Output mains
EN 55011 class A group 1
EN 61010-1
IEC1000-4-2,1000-4-3, 1000-4-4
External INPUT/OUTPUT signals (sockets)
Semi specifications connector
Output fan Voltage and Valve, forepump socket
RS 232/422/485 connection (optional)
Mains cable (3-wire, 3-meter long)
Pump cable (9-wire, 5-meter long)
19 kg (42.56 lbs)
70
TECHNICAL INFORMATION
Maximum 5 mA
Maximum 5 mA
87-900-929-01 (H)

Advertisement

Table of Contents
loading

This manual is also suitable for:

969-9448969-9449

Table of Contents