9.4.2
Replacing the incremental shaft encoder (potentiometer)
• Remove the cap on top of the incremental shaft encoder (see figure 42).
• With a socket wrench you can loosen the fastening nut.
Figure 42
• Loosen all electrical contacts to the shaft encoder on the back side of the front housing.
Figure 43
9.4.3
Replacing the display
• Loosen all electrical contacts to the display on the back of the front housing.
• Remove the 4 fastening screws of the display and pull away the display upwards.
Figure 44
© Hermle Labortechnik GmbH
SERVICE INSTRUCTIONS
remove
Service Manual Z 306
28