Glass Breakage; Service Life Of Rotors, Round And Rectangular Buckets, Accessories - Hermle Z 306 Service Manual

Universal centrifuge
Hide thumbs Also See for Z 306:
Table of Contents

Advertisement

but before the autoclaving. Residues of any cleaning liquids may cause fissures, whitening and
stains.
Gassterilization
Boxes, bottles and rotors may be gassterilized with Ethylenoxyd. According to the duration of the
application you may give long enough an airing to the items after the sterilization and before using
them again.
Attention: Because the temperature may rise during the sterilization, rotors, boxes and
bottles must not be closed respectively must be totally unscrewed
Chemical sterilization
Bottles, boxes and rotors may be treated with the usual liquid disinfectants.
Attention: Before applying any other cleaning resp. Decontamination method than
recommended by the manufacturer, contact the manufacturer to ensure that it will not
damage the unit or the rotor.

11.1.6 Glass breakage

With high g-values, the rate of glass tube breakage increases. Glass splinters have to be removed
immediately from rotor, buckets, adapters and the rotor chamber itself. Fine glass splinters will
scratch and therefore damage the protective surface coating of a rotor. If glass splinters remain in
the rotor chamber, fine metal dust will build up due to air circulation. This very fine, black metal dust
will extremely pollute the rotor chamber, the rotor, the buckets and the samples.
If necessary, replace the adapters, tubes and accessories to avoid further damages. Check the rotor
bores regulary for residues and damages.
Attention: Please check the relevant specifications of the tubes centrifuges with the
manufacturer!

11.1.7 Service life of rotors, round and rectangular buckets, accessories

Rotors and rotor lid made of aluminium or stainless steel, have a operating life of max. 7 years from
first use.
Transparent rotor lids and caps made of PC or PP as well as rotors, tube racks and adapters of PP
have a maximum service life up to 3 years from first use.
Tube racks, which are used in rotor no. 221.12 V03 have a service life up to 1 year from first use.
Condition for the operating life:
Proper use, damage-free condition, recommended care.
© Hermle Labortechnik GmbH
MAINTENANCE
Service Manual Z 306
44