Operating manual for
universal centrifuge
Z 306
manual.docx
Z306 Instruction
lish
© Hermle Labortechnik GmbH
05/2012

Advertisement

Table of Contents
loading

Summary of Contents for Hermle Labnet Z306

  • Page 1 Operating manual for universal centrifuge Z 306 manual.docx Z306 Instruction lish © Hermle Labortechnik GmbH 05/2012...
  • Page 3 © Hermle Labortechnik GmbH Z306 Instruction...
  • Page 5: Table Of Contents

    2.3.3  Preselection of brake intensity and acceleration ..................... 17  2.4  Radius correction..............................18  2.5  Program................................18  2.5.1  Storage of programs............................18  2.5.2  Recall of stored programs ..........................19  2.5.3  Leaving program mode............................. 20  manual.docx Instruction © Hermle Labortechnik GmbH Z306...
  • Page 6 Table 3: Max. speed and RCF-values for permissible rotors ...................VII  Table 4: Acceleration and deceleration times......................VIII  Table 5: Error messages ............................IX  Tabelle 6: Radius correction............................X  Table 7: Abbreviations used............................XII  Redemption form / Decontamination certificate.......................XIII  manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 7 Hermle centrifuges are intended exclusively for indoor use and for use by qualified personnel. Only Hermle original rotors and accessories might be used. Any other use or intended use is considered improper. From the resulting damage the company Hermle Labortechnik is not liable.
  • Page 8: Operating And Display Elements

    / Deceleration intensity time centrifugation time lid release quick short running start start centrifugation stop stop centrifugation prog calling stored programs store program store manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 9: Product Description

    „service“ „rcf“ h m s „accel“ 1.5 Signs and indications of the centrifuge 1.5.1 General Instructions for disposal (see chapter 6, P. 28) Direction of rotation – clockwise rotation for the rotor drive manual.docx © Hermle Labortechnik GmbH Z306 Instruction...
  • Page 10: Product-Nameplate (Example)

    PRODUCT DESCRIPTION Reference for loading rotors 1.5.2 Product-nameplate (Example) Company address: Hermle Labortechnik GmbH, Siemensstr. 25, D-78564 Wehingen TYPE: Type designation of the product REF: Order no. of the product Serial number of the product Date of manufacture MAX. Drehzahl: max.
  • Page 11: Warning And Information Signs

    020. • The HERMLE Z 306 is non explosion-proof and must therefore not be operated in explosion- endangered areas or locations. Centrifugation of flammable, explosive, radioactive, or such substances, which chemically react with high energy, is strictly prohibited. The final decision on the risks associated with the use of such substances is the responsibility of the user of the centrifuge.
  • Page 12: Following Rules Must Strictly Be Adhered To

    • Do not use accessories such as rotors and buckets, which are not exclusively approved by HERMLE Labortechnik GmbH, except commercially available centrifuge tubes made of glass or plastic. • Do not spin extremly corrosive substances, as they may cause material damages and impair mechanical resistance.
  • Page 13: Installation Of The Centrifuge

    (The socket for the power cord must be easy to reach respectively easy to disconnect). Switching it on using the mains power switch (I). Open the lid by using the button LID. • Remove the transport securing device of the motor. manual.docx © Hermle Labortechnik GmbH Z306 Instruction...
  • Page 14: Basic Adjustments

    All changed settings must be confirmed by the key „start“ (9). As an optical confirmation appears the word „store“ in the display „rpm/rcf“(A-1) - Only then the pre-selections are valid! (see figure 3) manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 15: Sound Signal Turn On / Off

    9 (loud) with the adjusting knob (1) (see figure 5). After you have stored the settings (see 1.7.1) you changed to the normal program mode again by switch off the centrifuge for a short while. Figure 5 manual.docx © Hermle Labortechnik GmbH Z306 Instruction...
  • Page 16: Song Selection For Sound Signal - End Of Run

    (1). (see figure 7). After you have stored the settings (see 1.7.1) you changed to the normal program mode again by switch off the centrifuge for a short while. Figure 7 manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 17: Call Up Operating Data (By Skilled Personnel Or Service Engineer Only!)

    (1). The respective error codes appear in the display „rpm/rcf“ (A-1). Please look up in „Table 5: error messages“ (see APPENDIX S. IX). Here as well you must shortly switch off the centrifuge for changing to the normal program mode again. Figure 8 manual.docx © Hermle Labortechnik GmbH Z306 Instruction...
  • Page 18: Operation

    On each rotor is designated how large the maximum load per hole is. (It is allowed to operate e.g a 12-place-rotor with 2 or 4 loaded tubes only. But the loaded borings must be opposite each other). manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 19: Loading Swing Out Rotors

    The bolts at the rotor must be greased with the HERMLE High EF oil (Order No. 34-5147). The sample tubes have to be filled evenly by eye and put into the drillings or tube racks The weight difference of the loaded buckets should not exceed approx.
  • Page 20: Loading And Overloading Of Rotors

    If In case of any questions, please contact the manufacturer! 2.1.5 Removing the rotor Untighten the rotor fixing nut complete (2. screw over the stiff point) and lift the rotor vertical out of the centrifuge. (See figure 10 and 11) manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 21: Lid

    All with number marked passages refer to figure 16 ATTENTION: Don't grip your fingers between lid and device or locking mechanism when closing the lid! manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 22: Preselection

    1. Range from 10 seconds up to 59 minutes 50 seconds in steps of 10 seconds 2. Range from 1 hour up to 99 hours 59 minutes in steps of 1 minutes 3. Range continuous run „cont”, which can be interupted by the key “stop” (10). manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 23: Preselection Of Brake Intensity And Acceleration

    All with number marked passages refer to figure 19 See „table 4: acceleration and deceleration times“ (APPENDIX P. VIII). There are shown the acceleration and deceleration times for the acceleration and deceleration stages 0 to 9 for permissible rotors. manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 24: Radius Correction

    Put the needed rotor into the centrifuge. By pressing the key „prog“ (11) in the display „time“ (A-3) appears the word „program“. With the adjusting knob (1) you can chose the desired program number. manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 25: Recall Of Stored Programs

    „rpm/rcf“ (A-1) flashes the word „rotor“ (M3). At the same time the word „FALSE“ and the stored rotor number „22x. xx“ (M4) will flashing by turns. All with number marked passages refer to figure 22. Figure 22 manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 26: Leaving Program Mode

    By the „stop“ key (10) (see figure 24) you can interrupt the run at any time. After pressing the key the centrifuge decelerates with the respective pre-selected intensity down to stand still. Figure 24 manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 27: Imbalance Detection

    Load the rotor as described in chapter 2.1.1 and 2.1.2. When inside the display “time” (A-3) the word “error” together with the number “02“ (see figure 25) appear, there could be following reasons: The imbalance switch is defective. Figure 25 manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 28: Maintenance

    (if available). The most suitable lubricant is the offered HERMLE High TEF oil – Order no.: 34-5147. Lubricants containing molycote and graphite are not allowed.
  • Page 29: Cleaning And Disinfection Of The Unit

    Before the autoclaving the PP-rotor and adapter must thoroughly be cleaned to avoid the burning in of dirty residues. You can disregard the consequences of some chemical residues to plastic materials at ambient temperatures. But at the high temperatures of the autoclaving those residues may corrode and manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 30: Glass Breakage

    Transparent rotor lids and caps made of PC or PP as well as rotors, tube racks and adapters of PP have a maximum service life up to 3 years from first use. Condition for the operating life: Proper use, damage-free condition, recommended care. manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 31 MAINTENANCE manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 32: Trouble Shooting

    At the right side of the centrifuge there is a plastic stopper (Figure 26). Remove this stopper, which is connected to the lid lock, horizontally from the housing until the centrifuge lid opens. • Now open the lid of the centrifuge. Figure 26 manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 33: Description Of The Error Message System

    4.2.2 Description of the error message system The error message „error“ (M11) is shown in the „time“ (A-3) display (see figure 27). Detailed information about possible error messages are in „table 5: error messages" (see Appendix P.IX). Figure 27 manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 34 TROUBLE SHOOTING manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 35: Receipt Of Centrifuges To Repair

    Decontamination certificate at goods return delivery (see APPENDIX P. XIV) We reserve the right to accept contaminated centrifuges. Further on all costs occurred for the cleaning and disinfection of the units will go to the debit of the customer’s account. manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 36: Transport, Storage And Disposal

    To document this they have been marked with the following identification: Because disposal regulations may differ from one country to another within the EU please contact your supplier if necessary. manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 37: Appendix

    Table 4: Acceleration and deceleration times ..............VIII   Table 5: Error messages...................... IX   Tabelle 6: Radius correction ....................X   Table 7: Abbreviations used ....................XII   Redemption form / Decontamination certificate ............. XIII   manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 38: Ec Conformity Declaration

    EG Konformitätserklärung EC Conformity Declaration Hermle Labortechnik GmbH - Siemensstr. 25 - D-78564 Wehingen – Germany Das bezeichnete Produkt entspricht den einschlägigen grundlegenden Anforderungen der aufgeführten EG-Richtlinien und Normen. Bei einer nicht mit uns abgestimmten Änderung des Produktes oder einer nicht bestimmungsgemäßen Anwendung verliert diese Erklärung ihre Gültigkeit.
  • Page 39 Write from operator Inventory-No.: Monitoring-Nr.: Environement: Maintenance contract: HERMLE Labortechnik GmbH or dealer service office Siemensstrasse 25 responsible service office 78564 Wehingen Tel.: (49)7426 / 96 22-17 Fax: (49)7426 / 96 22-49 responsible dealer manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 40 2 x 400 g 12000 min-1 221.17 V03 30 x 3,4 g 221.19 V02 4500 min-1 30 x 32 g 3250 min-1 221.24 V02 2 x 70 g 5000 min-1 221.25 V03 6 x 110 g manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 41 221.12 V03 4500 min-1 2716 x g 221.16 V03 12000 min-1 15133 xg 221.17 V03 4500 min-1 2830 xg 221.19 V02 221.24 V02 3250 min-1 1700 xg 5000 min-1 3801 xg 221.25 V03 manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 42 220.72 V06 220.87 V09 220.87 V10 220.96 V04 220.97 V04 221.12 V03 221.16 V03 221.17 V03 221.19 V02 221.24 V02 221.25 V03 in seconds Accelertaion time Decelertaion time ‐1 ‐1 from 0 min  ‐> U from U  ‐> 0 min VIII manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 43 Timeout between control unit and frequency converter Other error frequency converter Overspeed Rotor is not allowed in this centrifuge Inserted rotor does not exist in the program FALSE Rotor is not detected rotor no manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 44: Tabelle 6: Radius Correction

          220.96.05.04  With  9,7  ‐0,8        220.96.08.04           701.011  220.96.06.04  9,7  ‐0,8     701.012  220.96.07.04  8,3  ‐2,2  Angle rotor  603.000  220.97.04.04  9,8  0  220.97 V04  Swing out rotor        10,2  0  220.50 V08  manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 45 704.005 220.56.56.04 ‐1,1 12,5 Angle rotor           10,9 221.19.01 701.000 225.52.71.04 ‐4,5 701.010 220.96.08.04 11,3 ‐1,2 701.011 220.96.06.04 12,2 ‐0,3 10,6 701.012 220.96.07.04 10,5 ‐2,0 Swing outgrotor 221.24.03 14,36  221.24.01 Angle rotor         13,6 221.25.01 221.25.06 13,3 ‐0,3 manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 46 APPENDIX Table 7: Abbreviations used Symbol / Abbreviation Unit Description ‐1 U (=rpm) revolutions per minute [min RZB(=rcf) [x g] relative centrifugal force  ‐  Polypropylen  ‐  Polycarbonat accel  ‐  acceleration decel  ‐  deceleration manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 47: Redemption Form / Decontamination Certificate

    These substances have reached the equipment/assembly? …………………………..  Yes  No Which one, if yes: ________________________________________________________________ Description of the measures for the decontamination of the listed parts: I confirm the proper decontamination: Company/dept ._____________ place and date: ____________________________________ manual.docx XIII Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 48 APPENDIX Signature of the authorized person: ________________________________________ manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 49 NOTES manual.docx Z306 Instruction © Hermle Labortechnik GmbH...
  • Page 50 31 Mayfield Ave. Edison, NJ 08837 • USA • US • toll free: 888-LABNET1 • fax: 732 417-1750 • International • phone: +1-732-417-0700 • fax +1-732-417-1750 website: http://www.labnetinternational.com email: labnetinfo@corning.com Technical modification rights reserved. ©HERMLE Labortechnik GmbH 2012 manual.docx Z306 Instruction © Hermle Labortechnik GmbH...