Hermle Z 306 Service Manual page 37

Universal centrifuge
Hide thumbs Also See for Z 306:
Table of Contents

Advertisement

3) Speed and imbalance sensor (on the control board)
Figure 53
• Remove the areal as described in chapter 9.4.9.
• Remove the screws from the motor fixing.
• Lift the motor out of the unit.
Figure 54
• When reassembling the spare motor, take care of the wirings, as they must not be jammed.
• When tightening the motor mount screws, please pay attention the motor rubber mounts are not
being twisted.
• When connecting the motor with the electricity, take care of the rotating direction (Direction of
arrow) on the housing.
• Reassemble the unit in reverse order.
© Hermle Labortechnik GmbH
2
1
Service Manual Z 306
SERVICE INSTRUCTIONS