Application information

3.1 General

Prior to using the device on a patient, the user should become familiar with the
instructions for use and individual treatment methods to be used as well as the
indications / contraindications, warnings and application information. Additional
sources of information about the treatment should be followed.
These instructions for use must always be stored with the device and kept
accessible at all times for anyone authorised to operate this device.
After the device has been in transit on its side, place it upright and keep it in an
!
Caution
upright position for at least 30 minutes before switching it on. Otherwise the
compressor will be damaged.
Before use, ensure that the device is powered via a properly grounded mains socket
!
Caution
(electrical installation according to DIN VDE 0100 Part 710). The device must only
be operated with the supplied power cable. The power cable must be protected
against mechanical stress.
Magnetic and electrical fields can affect the function of the device. For this reason,
!
Caution
do not operate Cryo 6 in the vicinity of devices which generate strong
electromagnetic fields (X-ray or diathermy equipment, MRI machines). Please keep
a safe distance of several meters.
Cryo 6 is not suitable for use in areas with an explosive, flammable or combustive
!
Caution
atmosphere.
Do not place the device adjacent to heat sources (heating, hot mud products, sauna
!
Caution
etc.), and leave a gap of at least 50 cm between the device and a wall (to allow for
the supply of treatment products and cold air).
When used in combination with a laser device, the laser ventilation system must not
impede Cryo 6 cooling.
During use, the device is to be located in a position allowing direct access to the
!
Caution
device´s central mains supply so that it can be disconnected from the mains at any
time.
To avoid the risk of electric shock, the plug must be disconnected from the power
!
Caution
supply before performing any cleaning or maintenance activities.
Inspect the device before use. If there is any damage, it must not be used.
!
Caution
Only accessories provided by Zimmer MedizinSysteme GmbH must be used.
!
Caution
3