Technical Information - Zimmer Cryo 6 Instructions For Use Manual

Hide thumbs Also See for Cryo 6:
Table of Contents

Advertisement

Technical information

Mains power
Mains fuse
Protection class
Applied part
Dimensions
Weight
Operation
Transport
Storage
Note:
Evaporator temperature
generated
Air output temperature
(room temperature up to
25°C)
Max. treatment time that
can be set
Maximum load on glass
plate
100-120 V / 50 Hz / 60 Hz (9-12 A)
220-240 V / 50 Hz (7 A)
240 V / 60 Hz (7 A)
16 A circuit breaker in mains switch
I
Type B
H 645 mm x W 390 mm x D 680 mm
75 kg
+10°C to +35°C, 20% to 80% relative humidity, without condensation
at 700 hPa to 1060 hPa
-10°C to +50°C, 10% to 90% relative humidity, without condensation
at 600 hPa to 1060 hPa
0°C to +40°C, 10% to 90% relative humidity, without condensation
at 600 hPa to 1060 hPa
Storage and transport only in original packaging.
minimum (standby)
maximum (standby)
average:
- 25°C
at the start of treatment: - 31°C
maximum:
- 18°C (after 15 minutes of treatment)
Figures accurate to +/- 10%
99:59 min
The maximum weight and size of devices (such as laser devices) that can be
placed on the glass plate is 35 kg and 50 x 50 x 35cm (W x D x H).
Subject to technical changes.
115 V device
- 38°C
- 25°C
10
230 V device
- 43°C
- 25°C
Page 25

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents