Cleaning
Disinfection
- Before starting any maintenance and cleaning measures the device must always
be switched off at the main switch and the mains cable must be disconnected.
- Make sure that when cleaning and disinfecting the labelling of the device (such as
warnings, labels of control devices, identification plate) is not damaged.
- Make sure that during cleaning and disinfection no liquids penetrate the device. Do
not use sprays.
- If during cleaning or disinfecting liquid penetrates the device, please put the device
out of service, protect it from being used again and contact your service
representative.
- The device and its applied part are considered as uncritical in relation to hygiene
due to the use on non-injured and healthy skin.
Housing / accessories
Cleaning: In the event of visible contamination, the housing, cables and accessories
can be cleaned using commercially available alcohol-free plastic cleaners. Wipe the
surface until the dirt is removed, using a soft cloth soaked according to the
specifications of the manufacturer of the cleaning agent but not dripping wet.
Disinfection: We recommend that disinfection is to be carried out at least once a
week, as well as if there is any indication of contamination. Consult with your
hygiene specialist when doing so. Always perform cleaning prior to disinfection.
Housing, cables and accessories can be disinfected using disinfectant wipes. Use a
commercially available alcohol-free disinfectant for metal and plastic, with
bactericidal, virucidal and fungicidal properties. Observe the application instructions
of the manufacturer. Wipe all surfaces using a soft cloth soaked according to the
specifications of the manufacturer of the disinfectant, but not dripping, or with cloths
pre-impregnated with disinfectant (wipes).
Also observe requirements for drying or post-cleaning, where applicable.
Defrost water container Cleaning: The defrost water container (7) should be cleaned whenever it has been
emptied. Follow the procedure described under "Housing / Accessories".
Disinfection: Follow the procedure described under "Housing / Accessories".
!
Caution
The device may only be operated with the defrost water container in place.
Air filter
The air filter (9) should be cleaned regularly, and in any event after no more than
200 operating hours (maintenance notice appears on the display).
This is done by vacuuming the air filter from the outside using a commercially
available domestic vacuum cleaner.
Use the device only in a hygienic environment.
Note: