Manufacturer´s EMC declaration
Key performance features of the Cryo 6 are: smooth output of cold air, smooth operation of all functions.
Guidelines and manufacturer's declaration – Electromagnetic immunity
The Cryo 6 device is intended to be operated in the electromagnetic environment as indicated below. The customer or user
of the Cryo 6 device should ensure that it is used in such an environment.
Immunity
IEC 60601- test
tests
level
Conducted
3 V
Effective value
HF
150 KHz to 80 MHz
disturbances
according to
IEC 61000-
3 V/m
4-6
80 MHz to 2.5 GHz
Radiated HF
disturbances
according to
IEC 61000-
4-3
NOTE 1 At 80 Hz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not be applicable in all cases. The propagation of electromagnetic variables is influenced by
absorption and reflection from buildings, objects and people.
Compliance level
Electromagnetic environment - Guidelines
3 V
Portable and mobile radio equipment should not be used
Effective value
150 KHz to 80 MHz
at a distance away from the Cryo 6, including the cables,
that is less than the recommended safety distance which is
calculated according to the equation applicable to the
transmission frequency.
3 V/m
Recommended safety distance:
80 MHz to 2.5 GHz
with P as the rated output of the transmitter in watts (w)
according to the information of the manufacturer of the
transmitter and d as the recommended safety distance in
meters (m).
At all frequencies, the field strength of stationary radio
transmitter should be less than the compliance level
according to on-site testing
In the environment of devices which bear the following
symbols, interferences are possible: