Safety and maintenance

15.2 Maintenance

Cleaning the air filter
There is a filter element on the rear of the device (9) to ensure that the treatment air
and the cold air needed for the cooling machine are free from coarse dust particles.
It is normally sufficient to clean this regularly by vacuuming the whole of the outside
with a commercially available domestic vacuum cleaner, no later than when the
software issues a reminder after 200 operating hours. We recommend cleaning the
filter element more frequently in carpeted rooms or in areas where dust formation is
likely to be higher for other reasons.
Note:
The software message must be confirmed by pressing the "OK" key after the air filter
has been cleaned.
Emptying the defrost
When Cryo 6 is switched off after use, or when the defrosting program is started, the
water container
cooling system heats up and so produces defrost water.
The tank, which has a capacity of approximately 1 litre, can be removed by pulling it
out of the device and can then be emptied.
It should then be replaced in the device.
Please also follow the instructions for cleaning and disinfection (chapter 11).
15