Application Information

3
Before using the device on a patient, the user should become acquainted with the
operating instructions and individual treatment methods as well as the indications /
contraindications, warnings and application information. Additional sources of
information about types of therapy should be consulted.
Before use, ensure that the device is powered via a properly earthed plug with a
grounded outlet (electrical installation according to DIN VDE 0100 Part 710). The
device must only be operated with the supplied power cord. The power cord must be
protected against mechanical stress.
Operation of this equipment in the vicinity of strong electromagnetic fields (e.g.
tomographs, x-ray or diathermy equipment) may interfere with the operation of the
device. Please keep a safe distance of several meters.
clTrac is not suitable for use in areas with an explosive, flammable or combustive
environment.
To avoid the risk of electric shock; the device´s main plug must be disconnected
from the mains supply before performing any maintenance, service or reprocessing
activities.
Inspect the device before to use. If there is any damage, it must not be used.
Only a traction bank provided or recommended by Zimmer MedizinSysteme GmbH
must be used.
Only accessories provided or recommended by Zimmer MedizinSysteme GmbH
must be used.
WARNING: Use of accessories, transducers and cables other than those specified
or provided by the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this equipment
and result in improper operation
WARNING: Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30 cm (12
inches) to any part of the clTrac, including cables specified by the manufacturer.
Otherwise, degradation of the performance of this equipment could result
The clTrac must be placed on a level ground