Keysight N7744A User Manual page 68

N774 series optical multiport power meter
Hide thumbs Also See for N7744A:
Table of Contents

Advertisement

N O T E
Keysight N774x Series
If the tunable laser source module is rack- mounted the temperature and humidity
within the rack apply.
Polarization dependent responsivity (PDR)
Specifies the dependence of the measured power on the input state of
polarization (SOP) as ± half the difference between the highest and the
lowest measured power over all SOPs, expressed in dB.
Powers expressed in dBm.
Conditions: As specified.
Port separation (static)
Specifies the difference between the power P applied to one port and
the measured power D on any other port with no applied power (dark).
Excluding "Noise pp (dark)" (see,
"Drift (dark)" (see,
"Drift (dark)" on page
Powers expressed in dBm, port separation S expressed in dB,
S = P – D
Figure 44
Port separation (static)
Conditions: Applied power level as specified.
Port separation, dynamic
Specifies the difference between step power input P
or negative slope) applied to one port and the maximum measured
power change D
(positive or negative) on any other port without
chng
applied power (dark). Excluding "Noise pp (dark)" (see,
(dark)" on page
65) and "Drift (dark)" (see,
Specifications and Regulations Compliance
"Noise pp (dark)" on page
63).
"Drift (dark)" on page
4
65) and
(with positive
step
"Noise pp
63).
67