RHYTHMx AED ECG analysis algorithm

RHYTHMx AED ECG analysis algorithm
The RHYTHMx™ AED ECG analysis algorithm provides ECG
detection capabilities. The features available with the AED include the
following:
Detection Rate
Asystole Threshold
Noise Detection
Non-Committed Shock
Synchronized Shock
Pacemaker Pulse Rejection
SVT Discriminators
Supraventricular Tachycardia (SVT) Rate
Detection rate
All ventricular fibrillation (VF) and ventricular tachycardia (VT)
rhythms at or above this rate will be classified as shockable. All rhythms
below this rate will be classified as non-shockable. This rate is
programmable between 120 bpm (beats per minute) and 240 bpm via
MDLink Software by the Medical Director. The default Detection
Rate is 160 bpm.
Asystole threshold
The asystole baseline-to-peak threshold is set at 0.08 mV. ECG
rhythms at or below 0.08 mV will be classified as asystole and will not
be shockable.
Noise detection
The AED will detect noise artifacts in the ECG. Noise could be
introduced by excessive moving of the patient or electronic noise from
external sources like cellular and radiotelephones. When noise is
detected, the AED will issue the prompt "ANALYSIS
INTERRUPTED. STOP PATIENT MOTION" to warn the operator.
The AED will then proceed to reanalyze the rhythm and continue with
the rescue.
Powerheart® AED G3 Plus 9390A and 9390E
70-00914-01 F
2-3