162

Fine alignment of the purging flanges

Ensuring safety
Call up the
"Alignment" menu
Fine alignment of the
purging flanges
LS4000, LS4060 DIODE LASER ANALYZERS | OI/LS4000/O2-EN REV. A
No explosion protection is present when performing this work!
Risk of explosion.
Make sure that no explosive gases or dusts are present when
performing this work.
Proceed as follows:
Step
Procedure
1
Log on as an operator. The operator password is "xs2ls".
The "Main menu" is displayed.
2
Call up the "Alignment" menu (see page 164).
The "Alignment" menu appears.
The current values "Relative transmission" and "Absolute trans-
mission" are displayed.
3
Click on "Align".
The analyzer is now ready for fine alignment.
No.
Meaning
4 stud screws
1
4 fastening nuts
2