Insertion-Type Electromagnetic Probe Flowmeter
AquaProbe 2

2.7 AquaProbe Alignment

Warning. When inserting or removing the AquaProbe suitable restraining equipment must be used to
prevent the probe being forced out under pressure. Ensure that the valve is fully open.
Note. Safety restraint omitted for clarity.
Note. Measurement error due to misalignment (of <2) is <0.15%.
Fig. 2.15 Probe Alignment
16
Align parallel to pipe
(within 2°) see Notes
1
Slacken
See Notes
3
Tighten
40Nm
(30ft lbf)
2 Mechanical Installation
2
IM/AP
Issue 9