TOMOGRAPHIC EXPOSURE
15.3

Patient positioning with impression plate

B
A
90 Planmeca ProMax & Planmeca ProMax 3D s / 3D
Spread bite registration paste to the upper or lower side of
the impression plate.
Ask the patient to bite the plate so that the anterior teeth
coincide with the cross on the plate. The track of the
dental arch is formed in the registration paste.
Bite_this.eps
Mark the target area on the registration paste.
Place the impression plate in position. The tomographic
lights form a cross which indicates the imaging position.
Adjust the lights so that they hit the marking on the
registration paste as described in section 15.2 "Adjusting
position of target area" on page 88. Light A must be
parallel to tangent of the jaw curve and light B must be
perpendicular to the jaw.
User's Manual (2D)