PROGRAMS
Longitudinal tomography / transtomography programs
4x LNG Manual
Manual, 1-4 longitudinal images. The user can make manual
adjustments between exposures.
3x LNG Autoposition
Automatic, 3 longitudinal images. The exposure position is
automatically moved between exposures. The movement is
adjustable, and it can be 1, 3, 6, 9 or 12mm.
The selected target position is image no. 2.
3x LNG Automatic Angulation
Automatic, 3 longitudinal images with different angles. Images can
be used as stereopairs.
The C-arm angle is automatically changed between the exposures.
The angle of the first exposure is the selected angle +7°
counterclockwise, and the angle of the third exposure is the
selected angle +7° clockwise.
The selected target position is image no. 2.
24 Planmeca ProMax & Planmeca ProMax 3D s / 3D
1/3
3/3
2/3
1/3
3/3
2/3
3xLNG_Manual.eps
4xLNG_Auto_pos.eps
3xLNG_Auto_angle.eps