4.6 Function

4.6.1 Target vacuum

4.6.2 Gravity drainage mode

• On the main screen the target vacuum can be set directly to -5 mbar by pressing the
button.
• If the
symbol is touched during inactive therapy, the target vacuum is set to -5 mbar. To
Vac
5 5
start the therapy, it must be started manually by pressing the symbol
• If the
symbol is touched during active therapy, the target vacuum is set to -5 mbar.
Vac
5 5
NOTICE
) The change of the target vacuum is immediately effective. There is no confirmation necessary.
32
• Please note, an adjusted target vacuum over
-50 mbar may cause pain and injuries to the
patient.
• On the main screen the target vacuum can
be set directly by pressing the
Vac
buttons.
5 5
• ATTENTION: The change in the target
vacuum takes effect immediately. There is
no confirmation necessary.
• The target vacuum can be freely selected
between -5 and -100 mbar in steps of 1
mbar.
• If the buttons
and
permanently, the increase / decrease is
accelerated
• The target vacuum of -20 mbar is preset
when starting the device.
• The
button can be used to set the
Vac
5 5
target vacuum directly to -5 mbar.
For more information see "4.6.2
Schwerkraftdrainagemodus" on page 32
34.
If target vacuum is adjusted over -50 mbar the
notice appears ,,High target vacuum is set".
,
or
are pressed
Vac
5 5
.