19.2 Trace Measurement Problems and Solutions

 
 
Problem
 
 
 
 
 
Expected events are
not shown in the
trace
 
 
Special
events
appear at the end of
the trace
 
 
 

20 Warranty

 
 
 

20.1 Terms of Warranty

 
All AD INSTRUMENTS products are warranted against defective material and
workmanship for a period of one (1) year from the date of shipment to the original
customer. Any product found to be defective within the warranty period would be
repaired or replaced by AD Instruments free of charge.
 
 
Touch screen
needs calibration
 
 
Battery is not fully
charged.
Possible Reason
 
Events are too close
to each other.
 
SNR is poor
 
 
Wrong configuration
Wrong configuration,
e.g.
wrong
end
threshold setting.
The fiber under test
is too long to exceed
the range.
AD1000 User's Manual
 
Connect external USB mouse to
calibrate the stylus.
 
 
Re-calibrate the battery
Solution
u Try again with short pulsewidth
u Try testing from the other end of
the fiber
Try again with wide pulsewidth or
increase averaging time.
Check
OTDR
parameters,
Reflection threshold, non-reflection
threshold
and
re-analyze or test again.
 
 
Adjust configuration and test again.
 
 
Select proper range and test again.
like
end
threshold,
- 101 -