Adjusting Density from the Local Panel

Note
If your system includes the DryView V2 Keypad, do not use this Local Panel
procedure. Use the keypad procedure described in Section 4. Values set at the
keypad will override any values set in the Local Panel.
1. Press the Density key. The display will show DMAX = x.x. The number x.x is the
value of the maximum density desired.
2. Press the Test Print key. The imager will print a SMPTE test print.
3. Examine the test print. If the maximum density on the test print is not
acceptable, follow the steps below to adjust density.
4. Press the Density key.
5. Press the = key to increase density or the O key to decrease density. The display
will show the new density value. Repeat steps 2 and 3 until you reach an
acceptable maximum density.
Note
If you know the desired Dmax value, you can simply enter the value without
printing a test print. Follow the steps below.
1. Press the Density key.
2. Press the = or O key until the desired value is displayed.
3. Press the o (Enter) key to lock in the new value.
Density
Test
Print