Manually Closing the Film Cartridge

The imager automatically closes the film cartridge when the Door Open key is
pressed. If the imager is unable to close the cartridge, a P177 error displays.
Note
If the front door is opened when the film cartridge is open, the top sheets of film
in the cartridge are exposed. Make the room as dark as possible to minimize
exposure.
1. Press and hold the Open Door key on the Local Panel for 5 seconds to unlock
and open the front door.
2. Clear any films that may be jammed in the film cartridge area.
3. Turn the rollback knob (see Figure 5-8) counterclockwise to close the cartridge.
Figure 5-8. Manually Closing the Film Cartridge
2000 March Rev. B
8754020
Troubleshooting
Rollback
Knob
8100–54L