8.22. CYANURIC ACID
SPECIFICATIONS
Range
Resolution
Accuracy
Light Source
Method
REQUIRED REAGENTS
Code
HI93722‑0
REAGENT SETS
HI93722‑01
HI93722‑03
For other accessories see page 200.
MEASUREMENT PROCEDURE
• Select the
Cyanuric Acid
Method Selection section (see page 18).
• Fill the first cuvette with 10 mL of unreacted
sample (up to the mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show "‑0.0‑" when the meter is zeroed and ready for
measurement.
84
0 to 80 mg/L (as CYA)
1 mg/L
±1 mg/L ±15% of reading at 25 °C
LED with narrow band interference filter @ 525 nm
Adaptation of the turbidimetric method.
Description
Cyanuric Acid Reagent
Reagents for 100 tests
Reagents for 300 tests
method using the procedure described in the
Quantity
1 packet
# 1
10 mL
# 1