8.20. COPPER LOW RANGE
SPECIFICATIONS
Range
Resolution
Accuracy
Light Source
Method
REQUIRED REAGENTS
Code
HI95747‑0
REAGENT SETS
HI95747‑01
HI95747‑03
For other accessories see page 200.
MEASUREMENT PROCEDURE
• Select the
Copper LR
Method Selection section (see page 18).
• Fill the cuvette with 10 mL of unreacted sample (up to the mark)
and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show "‑0.0‑" when the meter is zeroed and ready for
measurement.
80
0.000 to 1.500 mg/L (as Cu)
0.001 mg/L
±0.010 mg/L ±5% of reading at 25 °C
LED with narrow band interference filter @ 575 nm
Adaptation of the EPA method.
Description
Copper Low Range Reagent
Reagents for 100 tests
Reagents for 300 tests
method using the procedure described in the
Quantity
1 packet
10 mL