Copper Low Range - Hanna Instruments HI83300 Instruction Manual

Multiparameter
Table of Contents

Advertisement

8.20. COPPER LOW RANGE

SPECIFICATIONS
Range
Resolution
Accuracy
Light Source
Method
REQUIRED REAGENTS
Code
HI95747‑0
REAGENT SETS
HI95747‑01
HI95747‑03
For other accessories see page 200.
MEASUREMENT PROCEDURE
• Select the
Copper LR
Method Selection section (see page 18).
• Fill the cuvette with 10 mL of unreacted sample (up to the mark)
and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show "‑0.0‑" when the meter is zeroed and ready for
measurement.
80
0.000 to 1.500 mg/L (as Cu)
0.001 mg/L
±0.010 mg/L ±5% of reading at 25 °C
LED with narrow band interference filter @ 575 nm
Adaptation of the EPA method.
Description
Copper Low Range Reagent
Reagents for 100 tests
Reagents for 300 tests
method using the procedure described in the
Quantity
1 packet
10 mL

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents