Turbidity and free/total chlorine meter
Hide thumbs Also See for HI 83414:

Advertisement

Quick Links

Instruction Manual
HI 83414
Turbidity and Free/Total Chlorine
Meter
w w w. h a n n a i n s t . c o m
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HI 83414 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Hanna Instruments HI 83414

  • Page 1 Instruction Manual HI 83414 Turbidity and Free/Total Chlorine Meter w w w. h a n n a i n s t . c o m...
  • Page 2: Warranty

    WARRANTY HI 83414 is guaranteed for two years against defects in workmanship and materials when used for its intended purpose and maintained according to instructions. Electrodes and probes are guaranteed for six months. This warranty is limited to repair or replacement free of charge.
  • Page 3: Table Of Contents

    TABLE OF CONTENTS WARRANTY .......................... 2 PRELIMINARY EXAMINATION....................4 GENERAL DESCRIPTION ......................5 ABBREVIATIONS ........................6 PRINCIPLE OF OPERATION ....................6 FUNCTIONAL DESCRIPTION ....................9 SPECIFICATIONS ......................... 11 GENERAL TIPS FOR AN ACCURATE MEASUREMENT ............... 12 START UP .......................... 20 RANGE SELECTION ......................
  • Page 4: Preliminary Examination

    Please examine this product carefully. Make sure that the instrument is not damaged. If any damage occurred during shipment, please notify your local Hanna Office. Each HI 83414 Bench Turbidity and Free/Total Chlorine instrument is supplied complete with: • Five Sample Cuvettes and Caps •...
  • Page 5: General Description

    The last calibration points, time and date can be checked. HI 83414 has a user-friendly interface with an easy to understand, graphic LCD. All messages are in plain text, easy to read and understand. Comprehensive contextual help is available at a simple key press.
  • Page 6: Abbreviations

    0 to 40 NTU range, using the nephelometric method. The HI 83414 instrument is designed to meet or exceed the criteria specified by the USEPA Method 180.1 and Standard Method 2130 B.
  • Page 7 The optical system of HI 83414 instrument is designed to have very low stray light, providing accurate results for low turbidity samples. However, special care must be taken when measuring low turbidities (see page 12 “General Tips for an Accurate Measurement”...
  • Page 8 Where: –log I/I = Absorbance (A) = intensity of incident light beam = intensity of light beam after absorption ε = molar extinction coefficient at wavelength λ λ = molar concentration of the substance = optical path through the substance Therefore, the concentration “c”...
  • Page 9: Functional Description

    FUNCTIONAL DESCRIPTION INSTRUMENT DESCRIPTION 1) Liquid Crystal Display (LCD). The LCD has backlight for better visibility in dark environments. 2) Keypad. Splash resistant. 3) Cuvette Lid. Close the cuvette lid prior to start a measurement. 4) Lamp lid 5) USB connector 6) Fuse holder 7) Mains switch 8) Mains connector...
  • Page 10 DISPLAY DESCRIPTION The display contains the following fields: 1) The current time in selected format 2) Information related to the measurement 3) Functional keys 4) Currently selected parameter 5) Measuring units 6) Measured value KEYBOARD DESCRIPTION The keyboard contains 8 direct keys and 3 functional keys with the following functions: The function of each of the three functional keys depends on the name displayed on the LCD above them.
  • Page 11: Specifications

    SPECIFICATIONS Turbidity Range - non ratio mode 0.00 to 9.99; 10.0 to 40.0 NTU 0.0 to 99.9; 100 to 268 Nephelos 0.00 to 9.80 EBC Resolution - non ratio mode 0.01; 0.1 NTU 0.1; 1 Nephelos 0.01 EBC Range - ratio mode 0.00 to 9.99;...
  • Page 12: General Tips For An Accurate Measurement

    GENERAL TIPS FOR AN ACCURATE MEASUREMENT HI 83414 is a highly accurate combined meter for turbidity and free & total chlorine. To meet the instrument’s performance and fully benefit from its features, it is very important for the analyst to use proper measurement techniques for accurate, precise and repeatable readings.
  • Page 13 Note: In colorimetric measurements, when it is possible, use the same cuvette both for zeroing and measurement. If this is not possible, always match the cuvettes. In turbidimetric measurements, if you are using multiple cuvettes, always match the cuvettes. CUVETTE HANDLING The cuvettes should be free of scratches or cracks.
  • Page 14 Note: The supplied cloth for oiling should be stored together with the silicone oil bottle and cuvettes, taking care to avoid contamination with dirt. After a few oiling procedures, the cloth will contain enough oil to wipe the bottle with it without adding more oil. From time to time add some drops of oil on the cuvette to provide the necessary oil quantity in the cloth.
  • Page 15 • Insert the cuvette into the instrument and press the Read key. Record the reading. • Open the instrument lid, slightly rotate the cuvette and take a new reading. • Repeat the last step until you read the lowest NTU value. •...
  • Page 16 • Turn the instrument ON. • Insert the first cuvette into the instrument and press the Read key. Record the reading. • Mark this position on the thicker white band on the top of the cuvette with a water resistant pencil.
  • Page 17 • Alternatively, keep the Read key pressed and, after the first value is displayed, open the lid and start rotating the cuvette until the read value matches the first cuvette. • Mark this position on the second cuvette with a water resistant pencil. •...
  • Page 18 ADDITION OF SURFACTANT A surfactant addition works by changing the surface tension of the water. In this way bubbles are released from the sample. This method is effective in samples that are supersaturated with air. The procedure consists in the addition of a drop of surfactant in the cuvette before adding the sample to be analyzed.
  • Page 19 ADDING REAGENT (COLORIMETRY only) • Because the reagent quantity is set up to react with 10 mL of sample, it is very important to fill the cuvette correctly. The liquid in the cuvette forms a meniscus; the bottom of the meniscus must be at the same level with as the 10 mL mark.
  • Page 20: Start Up

    START UP The HI 83414 Turbidity / Free & Total Chlorine benchmeter is supplied with all necessary accesories for making measurements. Unpack the instrument and place it on a flat surface. Do not place the instrument under direct sunlight. Connect the instrument to the mains with the provided power cord. Pay attention that the mains voltage match the value printed on the back of the instrument.
  • Page 21: Tutorial Mode

    HELP MODE The HI 83414 offers an interactive contextual help mode that assists the user at any time. To access help screens, just press HELP. Depending which menu you are in, a screen with additional information will appear.
  • Page 22: Measurement Procedure For Ratio/Non Ratio Turbidity Range

    MEASUREMENT PROCEDURE FOR RATIO / NON RATIO TURBIDITY RANGE When taking turbidity measurements, several basic rules should be followed: • Never use cuvettes with scratches or cracks because they can cause inaccurate readings. • Always cap the cuvettes to avoid spillage of the sample into the instrument. •...
  • Page 23 NORMAL MEASUREMENT This type of measurement is best suited for regular readings, when the sample is stable and normal accuracy is required. In normal mode, the measurement takes about 10 seconds and the lamp is ON for a minimum period of time (about 7 seconds). •...
  • Page 24 HI 83414 automatically selects the correct turbidity range to display the results with the highest accuracy. If the measured value is higher than 4000 NTU (980 EBC or 26800 Nephelos), the display will show the maximum value blinking and the message “Out of range”...
  • Page 25: Calibration Procedure For Ratio/Non Ratio Turbidity Range

    HI 83414 turbidimeter is supplied with 5 AMCO standards: <0.1 NTU, 15 NTU, 100, 750 NTU and 2000 NTU. The Hanna standards are specially designed for this instrument. Turbidity standards have a shelf life and should not be used after the expiration date.
  • Page 26 To enter calibration, press the CAL CHECK key while in main screen. The first screen of GLP information is displayed. Press the Cal key to start calibration. It is possible to interupt calibration procedure at any time by pressing the CAL CHECK key . TWO-POINT CALIBRATION (Ratio and Non Ratio Turbidity) •...
  • Page 27 • Close the lid and press the Read key. The display will show the value blinking and the lamp icon during measurement. • If Non ratio Turbidity range is selected, the display will briefly show “Store...” and the two point calibration is saved. The instrument returns in the main screen.
  • Page 28 • Close the lid and press the Read key. The display will show the value blinking and the lamp icon during measurement. • At the end of the measurement, the fifth proposed calibration point (2000 NTU) is displayed. • At this time it is possible to exit calibration by pressing the End key.
  • Page 29: Measurement Procedure For Free/Total Chlorine

    OUT OF CAL RANGE FUNCTION The instrument has a mechanism to prevent taking measurements in a range where the calibration does not assure the best results. The message “Out of Calibration Range” is displayed on the message line in the following situations: •...
  • Page 30 • Place the cuvette into the instrument. Align the mark on the cuvette with the sign on the instrument top and close the lid. • Press the Zero key. The dashes and the lamp icon will blink on the LCD, depending on the measurement phase.
  • Page 31 • Press the Timer key. The display will show the stopwatch icon and the countdown prior to measurement. Alternatively wait for: Free Chlorine Total Chlorine 1 minute or 2 minutes and 30 seconds and then just press the Read key. Dashes and the lamp icon will blink on the LCD, depending on the measurement phase.
  • Page 32: Cal Check Procedure For Free/Total Chlorine

    2 minutes after adding the powder reagent. CAL CHECK PROCEDURE FOR FREE/TOTAL CHLORINE The HI 83414 free and total chlorine colorimeter has a powerful CAL CHECK function that allows the user to check the instrument calibration against a NIST traceable standard before making a set of measurements.
  • Page 33 • Place the zero cuvette (A) in the instrument with the mark aligned with the mark on the instrument top. • Close the lid • Press the Zero key. The dashes and the lamp icon will blink on the LCD, depending on the measurement phase.
  • Page 34: Calibration Procedure For Free/Total Chlorine

    CALIBRATION PROCEDURE FOR FREE/TOTAL CHLORINE Use the provided standard solution to calibrate the free or total chlorine range of the HI 83414. Do not calibrate the instrument with standard solutions other than Hanna CAL CHECK™ Standards, otherwise erroneous results will be obtained. For accurate calibration please perform test at room temperature, 18 to 25 ºC (64.5 to 77.0 ºF).
  • Page 35: Restore Factory Calibration

    CALIBRATION ERROR MESSAGES • The calibration is successfully performed if the reading is in certain limits. If the CAL CHECK™ standard value is too high, the display will show a high standard message. If this message appears, check if the correct cuvette was used. •...
  • Page 36: Good Laboratory Practice (Glp)

    GOOD LABORATORY PRACTICE (GLP) The HI 83414 has built in complete GLP information. The calibration date and the calibration points are displayed in a comprehensive mode for each range. To display the GLP information, simply press CAL CHECK key. A screen with instrument serial number and with information about the calibration is displayed.
  • Page 37: Log And Log Recall

    LOG AND LOG RECALL The HI 83414 has a powerful log function that could store up to 200 records. Each record contains: • the measuring range, • the reading value, • the measuring units, • the date and time of the measurement, •...
  • Page 38: Setup

    SETUP In the Setup mode, the instrument’s parameters can be changed. Some parameters affect the measuring sequence and others are general parameters that change the behaviour or appearence of the instrument. The items that are related to the selected parameter do not appear in the setup when another parameter is selected.
  • Page 39 Date / Time This option is used to set the instrument’s date and time. Press the Modify key to change the date/time. Press the ⊳ functional keys to highlight the value to be modified (year, month, day, hour, minute or second). Use the keys to change the value.
  • Page 40 Tutorial Option: Enabled or Disabled. If enabled this option will provide the user short guides on the screen. Press the functional key to enable/disable the tutorial mode. Beeper Option: Enabled or Disabled. When enabled, a short beep is heard every time a key is pressed.
  • Page 41: Lamp Replacement

    LAMP REPLACEMENT The instrument tungsten lamp has a life longer than 100,000 measurements. In case of lamp failure, the defective lamp can be easily replaced. When the lamp is broken, the instrument displays a lamp error message. To replace the lamp follow the next steps: •...
  • Page 42: Pc Interface

    Silicon oil (15 mL) HI 93711-01 Reagents for 100 Total Chlorine tests HI 93711-03 Reagents for 300 Total Chlorine tests HI 83414-11 Calibration set for turbidimeter(<0.1, 15, 100 750 and 2000 NTU) OTHER ACCESSORIES HI 731318 Cloth for wiping cuvettes (4 pcs.) HI 731331 Glass cuvettes (4 pcs.)
  • Page 43 To avoid damage or burns, do not put the instrument in microwave ovens. For your own and the instrument safety do not use or store the instrument in hazardous environments. Hanna Instruments reserves the right to modify the design, construction or appearance of its products without advance notice.